Safety of Mass Measurement in Pharmaceutical Industry. Basic Metrological Tests (part III)

Safety of Mass Measurement in Pharmaceutical Industry. Basic Metrological Tests (part III)

Specific requirements of this area are a result of the fact that the weighing devices are one of many components used in the process of production of pharmaceuticals. They may influence medicine quality and as a result the patient's health. Apart from regulations obligatory within a particular area, the pharmaceutical industry conforms to guidelines provided in so called pharmacopoeia. Pharmacopeia is a book of a continental scope, there are for example American, Japan, European, Russian pharmacopoeias.

 

Requirements concerning the weighing instruments are comprised in two USP sections:

  • General Chapters, Apparatus for Tests and Assays <41 Balances>,
  • General Information, <1251 Weighing on an Analytical Balance.

Section 41 provides the following requirement for the repeatability and accuracy:

  1. Repeatability is satisfactory if two times the standard deviation of the weighed value, divided by the nominal value of the weight used, does not exceed 0.10%. If the standard deviation obtained is less than 0.41d, where d is the scale interval, replace this standard deviation with 0.41d,
  2. The accuracy of a balance is satisfactory if its weighing value, when tested with a suitable weight(s), is within 0.10 % of the test weight value. A test weight is suitable if it has a mass between 5% and 100 % of the balance's capacity.

Section 1251 is not an obligatory one however it explains and defines terms crucial when it comes to electronic balances. Compliance with the pharmaceutical industry requirements means conformity with the provisions comprised in section 41 – mainly concerning indication repeatability. Repeatability is the most important balance parameter deciding about the measurement accuracy. This parameter is used to determine so called minimum mass (Minimum Sample Weight), which specifies the weighing range start. Weighing samples of mass value lower than the minimum sample weight value is disallowed.

USP 41 requirements regarding the repeatability of the minimum weight have been adopted by the European Pharmacopeia, General Chapter 2.1.7 Balances for Analytical Purposes. They will be obligatory starting from January 2022.  

 

Minimum Weight – Measurement Precision

SOP

MSW Determination

Definition

USP 41, Ph. Eur. 1.2.7.

Equipment

Mass standard of weight value lower than 5% of the maximum capacity of the balance. The test may be performed using a mass standard of weight value slightly greater than the expected MSW value – it is important to apply mass standard that can be deposited onto the weighing pan shock-free. 

Method

Manual

Load the weighing pan 10 times with a mass standard of a respective nominal value and record the weighing results. Balance indications for an unloaded weighing pan may be zero prior to and between the measurements. Calculate standard deviation (S) for the measurement series. Balance adjustment is not required prior to the test start.

Limits

m – the lowest weight value of a mass standard, which corresponds with the to-be-weighed substance quantity

S – standard deviation for the series of 10 measurements performed using the mass standard

Standard deviation (S) is used for determination of the MSW value by the following equation: MSW = 2000 · S.

Interpretation

The smallest standard deviation is 0.41d, therefore the smallest MSW value, depending on the reading unit (d), may be:

 

d = 1 mg
MSW = 0.41 · 1mg · 2000 = 820 mg (PS 1000.X2)

d = 0.1 mg
MSW = 0.41 · 0.1 mg · 2000 = 82 mg (AS 220.R2 PLUS)

d = 0.01 mg
MSW = 0.41 · 0.01 mg · 2000 = 8.2 mg (XA 52.4Y PLUS)

d = 0.001 mg
MSW = 0.41 · 0.001 mg · 2000 = 0.82 mg (MYA 5.4Y PLUS)

d = 0.0001 mg
MSW = 0.41 · 0.0001 mg · 2000 = 0.082 mg (UYA 2.4Y PLUS)

 

 

 

In practice it is recommended to apply slightly greater MSW values than those calculated on the basis of tests. This is mainly due to the fact that the repeatability depends on ambient conditions, good operator skills, etc. Certain repeatability variation shall be expected, it should be accounted for in risk analysis for mass measurement processes. 

Remark

Weight value of the mass standard used in the course of the test shall not be greater than 5% of the Maximum capacity. If the expected MSW is 40 mg, then the test weight may be, for example, 100 mg or 200 mg mass standard. While selecting a mass standard it is necessary to take into account the ergonomics when it comes to loading, no shocks.

 

 

Accuracy – Measurement Precision

SOP

Weighing process accuracy

Definition

USP 41, Ph. Eur. 1.2.7.

Equipment

Mass standards of nominals ranging between 5÷100% of the maximum capacity.

Method

Manual

  • adjust the balance,
  • zero the indication,
  • load the weighing pan centre with the mass standard,
  • write down the weighing result,
  • check if the following condition is met:

Safety of Mass Measurement in Pharmaceutical Industry. Basic Metrological Tests (part III)

Where:

m – mass standard weight value
I – indication of the balance for the mass standard 

Limits

The absolute value of the relative error may be greater than 0.0005.

Interpretation

A highly significant issue in the course of tests is the weight error related to the accuracy class (E2, F1, F2 etc.). When carrying out the tests with use of mass standards, it is necessary to account for both the weight value given on the calibration certificate, and the mass determination uncertainty.

Remark

Measurement accuracy depends on the balance sensitivity, deviation of linearity and the error of eccentricity. The most important of these three factors is balance sensitivity, regulated by means of the automatic balance adjustment (Figure 5). The influence of the remaining two, i.e., the error of eccentricity (loading of the weighing pan centre) and the deviation of linearity is insignificant. Irrespective of the above it is recommended to take the limit of 0.05 % for each.    

 

Adjustment – Adjustment Report

Adjustment is a process consisting in introduction of corrections due to which, regardless of the load, the balance indications are correct. In most professional laboratory solutions, just like in the case of Radwag devices, the adjustment is an automatic operation. It is triggered based on time, temperature change and operator's schedule with a to-do-list. After each adjustment a report is generated, the report features basic information concerning the process. Report example is shown in figure 22. These days more often the traditional printout is replaced with an electronic version either stored in the balance database or archived in other form. 

 

Autotest GLP           

Autotest GLP serves to determine mass measurement precision. It is taken for granted that if the measuring system is valid then multiple weighing of the same (adjustment) mass shall give similar results (low standard deviation value). The procedure may be repeated many times, wherein the obtained values must be registered. It is however necessary to always remember that mass measurement of objects other than mass standard does not have to be precise enough. If precision obtained as a result of the Autotest function operation is excellent and weighing of other loads gives much worse result, then it is necessary to correct the weighing methodology since the measuring system is efficient for sure. Autotest GLP report has been presented in figure 14.       

                       

Autotest Filter

Operation of  the Autotest Filter function consists in indicating an optimal combination of the measurement signal filtering method and the value release criterion. Most Radwag laboratory balances offer 5 signal filtering levels and 3 value release options. It is a very difficult task to check manually all these options and combinations, and in practice, in the laboratory, it is almost impossible (time and costs). With regard to the above a special function has been designed, it automatically checks both precision and weighing time for each combination of signal filtering and value release options. Finally, a report with the best possible solution for time and precision is generated.

Table 6. Example of the Autotest function report (brief version)

 

----------- Autotest FILTER Report -----------

 

Balance type                            

XA 4Y 

 

Balance ID                               

876573 

 

Operator                                 

Admin

 

Application revision                

L1.4.15 K

 

Date                                         

2021.05.27 

 

Time                                        

10:14:23 

 

------------------------------------------------------------------------

 

Reading unit                            

Internal weight mass                          

Temperature: Start               

Temperature: Stop                  

Humidity: Start                                 

Humidity: Stop                                 

0.0001 g

209.65432 g 

24.27 °C  

24.39 °C

51%

53%

 

 

 

 

 

 

 

 

 

Filter                            

Value release             

Repeatability                

Stabilization time               

Fast                         

Fast                           

0.00007 g                             

1.688 s

 

Filter                            

Value release             

Repeatability                

Stabilization time             

Fast                             

Fast and reliable                            

0.00007 g                             

2.255 s

 

Filter                            

Value release             

Repeatability                

Stabilization time              

Fast                             

Reliable                             

0.00007 g                             

2.760 s

 

 

Filter                            

Value release             

Repeatability                

Stabilization time              

Normal                      

Fast                             

0.00007 g                             

1.894 s

 

 

Filter                            

Value release             

Repeatability                

Stabilization time             

Normal                      

Fast and reliable                            

0.00004 g                             

2.423 s

 

 

Filter                            

Value release             

Repeatability                

Stabilization time             

Slow                           

Fast and reliable                                                      

0.00002 g                             

5.833 s

 

 

Where:

 

The shortest measurement time

 

The best measurement precision

 

The smallest product value of time and standard deviation

 

Current settings

 

We do recommend reading previous and next parts of the article titled Safety of Mass Measurement in Pharmaceutical Industry. Basic Metrological Tests:

Get familiar with RADWAG product range intended for the pharmaceutical industry: 

Safety of Mass Measurement in Pharmaceutical Industry. Basic Metrological Tests (part III)

Newsletter

Subscribe to the newsletter if you would like to receive RADWAG news.

View more
Effective Laboratory Management. How Does Process Automation Improve Efficiency?
2026-05-12

Effective Laboratory Management. How Does Process Automation Improve Efficiency?

The growing number of tests, the need to maintain the highest quality of results, and the limited availability of qualified staff mean that laboratory management today requires not only organisational skills but also the informed use of modern technologies.
More  ➜
Moisture Analyzer as a Quality Control Tool in Plastics Production – Accuracy and Effectiveness of RADWAG Technology
2026-04-28

Moisture Analyzer as a Quality Control Tool in Plastics Production – Accuracy and Effectiveness of RADWAG Technology

In modern plastics production, precise control of granule moisture content determines the stability of the technological process, the strength of parts and the minimisation of material losses. Read how RADWAG’s technological solutions allow your company to achieve full repeatability and process safety.
More  ➜
How Much Does a Lab Technician’s Time Cost You? The Hidden Costs of Manual Calibration
2026-04-14

How Much Does a Lab Technician’s Time Cost You? The Hidden Costs of Manual Calibration

Although manual pipette calibration is still widely used, it significantly increases working time and generates hidden costs which often go unnoticed in day-to-day laboratory operations. Fortunately, there are modern solutions available which can boost laboratory productivity.
More  ➜
Pallet Scale vs. Beam Scale – a Complete Guide to Industrial Scales
2026-03-31

Pallet Scale vs. Beam Scale – a Complete Guide to Industrial Scales

How does a U-type pallet scale differ from a beam scale? Are beam scales suitable for weighing machines and tanks? Why is it worth buying a pallet scale directly from the manufacturer? You will find answers to these and other questions in this article.
More  ➜
What Does ‘Precision Balances’ Mean?
2026-03-17

What Does ‘Precision Balances’ Mean?

Are precision balances inherently more precise than other balances? After all, the range of electronic balances includes laboratory balances, which are even more accurate, so where does this name come from?
More  ➜
Professional Weighing Workstations, Antivibration Tables and Weighing Tables by RADWAG – Elimination of Vibrations in the Laboratory
2026-03-03

Professional Weighing Workstations, Antivibration Tables and Weighing Tables by RADWAG – Elimination of Vibrations in the Laboratory

A weighing table is designed to eliminate ground vibrations which interfere with the operation of laboratory equipment, especially balances. Which balances and scales should weighing tables be used with? What is the difference between a weighing table, an antivibration table and a weighing workstation? You will find answers to all your questions in this article.
More  ➜
How to Choose a Scale for Use in Harsh Industrial Environments?
2026-02-17

How to Choose a Scale for Use in Harsh Industrial Environments?

In many industries, precise mass measurement must be performed in conditions far from laboratory conditions. Which RADWAG solutions have been designed specifically for such applications?
More  ➜
Balance as a Quality Guardian. Net Weight Control in Small-Scale Food Production
2026-02-03

Balance as a Quality Guardian. Net Weight Control in Small-Scale Food Production

The net weight declared on the label is a promise to the customer – information about how much product is actually in the package. In the food industry, keeping this promise is the foundation of credibility.
More  ➜
How to Teach Accurate Weighing? Analytical and Precision Balances in University Laboratories
2026-01-20

How to Teach Accurate Weighing? Analytical and Precision Balances in University Laboratories

At RADWAG, we have been observing for years how important our devices are in educating future generations of specialists. In this article, we suggest how to effectively teach weighing and select the optimal balance for a teaching laboratory.
More  ➜
Summary of 2025
2026-01-07

Summary of 2025

The 40th anniversary of the company, victory in a prestigious industry competition, new products – read about what RADWAG has been up to over the past year.
More  ➜
Best Wishes for Christmas and New Year
2025-12-23

Best Wishes for Christmas and New Year

Merry Christmas and a happy New Year to all RADWAG employees, partners and friends.
More  ➜
How To Choose a Weighing Device for Checking Raw Materials in the Cosmetics Industry?
2025-12-09

How To Choose a Weighing Device for Checking Raw Materials in the Cosmetics Industry?

In the cosmetics industry, precise weighing of ingredients determines the quality and safety of products, which is why choosing the right balance or scale for the laboratory and quality control department is half the battle.
More  ➜
Choosing a RADWAG Moisture Analyzer – Which One Will Work Best in Your Laboratory?
2025-11-25

Choosing a RADWAG Moisture Analyzer – Which One Will Work Best in Your Laboratory?

A moisture analyzer is an essential device in many laboratories, as it allows you to quickly and accurately determine the moisture content of samples. MA X7.A, PMV 5Y, MA 5Y or MA R – which RADWAG moisture analyzer should you choose and why?
More  ➜
« Back

© 2026 RADWAG Balances and Scales.

All rights reserved.